Verification Solutions

Zero Doubt, Absolute Proof: Guarantee the Accuracy of Your Test

You may have the world’s best leak testing device, but there is one critical question that regulations and quality auditors will always ask: “How do you prove that your device is measuring accurately at that moment?”

Adlema Verification Solutions take the measurement sensitivity and repeatability of your test device to the highest level. With Leak Masters and Self Check Modules integrated into your system, eliminate operator initiative and guarantee the accuracy of every test according to international standards.

GBH Series - Leak Masters

The most reliable way to make sure everything is running correctly on your production line is to test the system with a known fault. GBH Series Leak Masters are specially designed for the daily / periodic verification of your leak testing device and installed system.

These masters can be easily connected to the port on the device and transform a 100% approved / OK product from your production line into an artificially created sample with a very small, known leak. This allows you to test whether your system can instantly detect that leak.

Real-Time “Virtual Leak”

It instantly converts an approved product from your production line into a leaking sample by integrating it into your system. This allows you to live-test your device’s current response time and leak detection capability under real-world conditions.

Wide Pressure Scale

Whether your system operates under deep vacuum at -0.5 Bar or in high industrial pressures such as 20 Bar, your device verification is configured entirely according to your production parameters.

International Traceability and Regulatory Compliance

Prove that your leak tests comply with ISO, FDA and GMP standards using traceable data. Adlema Leak Masters can be delivered with an “Additional Calibration Certificate” upon request for your audit processes.

PRESSURELeak Rate for Air at 20°C and 1 ATM Pressure (sccm)
457910111315172035
500 mBarxxx0.520.650.781.372.543.274.5313.87
750 mBarx0.370.751.191.481.792.53.334.285.9218.14
1 Bar0.270.420.831.381.712.062.883.844.946.8320.94
2 Bar0.410.641.252.072.563.14.335.777.4110.2531.42
4 Bar0.681.062.093.464.275.177.229.6112.3517.0952.36
6 Bar0.971.4962.934.845.987.2410.1113.4617.2923.9373.31
10 Bar1.52.354.67.619.411.3715.8921.1627.1737.61115
20 Bar2.874.488.7914.5417.9521.7230.344051.8971.81219
-0.1 Barxxx0.40.50.610.851.131.452.016.15
-0.2 Barxxx0.540.670.811.131.51.942.688.2
-0.5 Bar0.130.210.410.690.851.031.441.922.473.4110.47

Verifiable Sensitivity, Measurable Performance

SL Series - Self Check / Automatic Verification Module

Leaving the quality verification process to operator initiative creates the risk of a false OK decision in high-volume production lines. The SL Series Self Check Module is an autonomous verification unit that automatically subjects your leak testing device to performance and fault checks at the intervals you define. The module starts a trial test through the selected leak master, confirms the accuracy of the sensor, and physically guarantees that your device continues to operate with its original measurement stability.

Customizable Automatic Verification

Your leak testing device is autonomously subjected to performance and fault checks at the exact intervals you define, without human intervention. It eliminates the risk of operators skipping verification during production intensity.

Original Sensor Sensitivity

When the defined verification period arrives, the system automatically activates and tests the sensor accuracy through the selected leak master. This physical verification confirms that the device still maintains its original measurement performance and calibration stability.

Brand-Independent System Integration

High-level automation reliability is not limited only to the Adlema ecosystem. It can also be easily connected and used with other leak testing devices already present on your production line.

Perfect
Your System

Crown the daily operational verification you perform on your production line with GBH Leak Masters and SL Self Check Modules with internationally recognized official documentation. Rely on Adlema’s Accredited Calibration Laboratory for the periodic pressure calibration of your leak testing devices and all pneumatic measuring instruments in accordance with ISO and GMP standards.

Keep your devices audit-ready at all times and your measurements within zero-defect tolerance.

Privacy Overview

This website uses cookies so that we can provide you with the best user experience possible. Cookie information is stored in your browser and performs functions such as recognising you when you return to our website and helping our team to understand which sections of the website you find most interesting and useful.